This "small/BIG-SCALE" MultiMeter is a "Classic": No large central switch to
change ranges - instead it has a thumb-wheel on the right side and a vertical
"Radio Scale" ( with rope & winch ) on the right side. The A-Ω-V Scales cover
the whole front of the 140 x 105 x 50 mm3 ( 5_1/2" x 4" x 2" ) instrument.
M U L T I M E T E R » MX-202 « by ITT-METRIX, Paris / France (1975)
******************************************************************************
Class 1.5 (DC) & 2.5 (AC) ; 40 kΩ / Volt (DC) & 1 kΩ / Volt (AC)
< Ri(I) = 2 kΩ >
50 mV .. 500 V ; 25 μA .. 0.5 A & 1000 V / 5 A DC Ranges
15 V .. 500 V ; 50 μA .. 0.5 A & 1000 V / 5 A AC Ranges
(x1) .. 20kΩ , (x10) .. 200kΩ , (x100) .. 2MΩ ;
w. 1.35 V SPECIAL BATTERY »RM-1R« MALLORY Cell
Because of its stability, NO ZERO-Ω-CALIBRATION is necessary!
0 dB .. +25 dB ( as Difference Measurement )
(1.) The FUSE-"Repair" is quick & easy ( see DETAIL Picture on the left ) ...
(1.) (2.)
The ORANGE SQUARE covers a little SPOOL with a very thin 0.05 mm Ø SILVER-WIRE.
Pull it out of the SLOT and mount it between the 2 SCREWS over the "WINDOW".
(2.) A T T E N T I O N :
-------------------------
DC & AC uses the same plugs for "Normal Ranges" - only the "Extended Ranges"
uses the + 1000 V (DC)/(AC) or + 5 A (DC)/(AC) Plugs - The SELECTOR has to be
set either on DC or AC-Range. As a reminder, I put COLOR-MARKS beside the plugs
BLUE / RED, red, red ; YELLOW / YELLOW, yellow, yellow. ( See right picture )
In these cases there is N O F U S E PROTECTION as shown in the DIAGRAM ...
|